5 kft ipol - titration parameters, Titration parameters – Metrohm tiamo 2.1 Manual User Manual

Page 753

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

739

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Titrando
Range

1 ... 999 µL/min

Default value

10 µL/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is only
shown for Stop criterion = time. The selection inf. stands for infinite.

Titrando, 855
Range

0 ... 999 s

Default value

10 s

Titrino
Range

0 ... 999 s

Default value

10 s

Selection

inf.

Stop time

The titration is stopped if the absolute stop time defined here has elapsed
since the start of the titration. This parameter is only shown for Stop cri-
terion
= time and Delay time = inf.

Titrino
Range

1 ... 999999 s

5.6.3.5.2.5

KFT Ipol - Titration parameters

Tab: Method

KFT Ipol Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.

Titration direction

Selection of the titration direction. If two endpoints are set, then the titration
direction is already established and the following selection of titration
direction
is ignored.