5 Method
tiamo 2.1
■■■■■■■■
1071
device, it will automatically be reduced during dosing to the largest possible
value.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The filling rate should be reduced for viscous liquids.
5.6.6.3.5.5
DOS U - Monitoring
Tab: Methods
▶ DOS U ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Input
25 characters
The following monitoring processes can be switched on and defined on this
tab:
Monitoring Measured value
Monitoring Measured value
on | off (Default value: off)
If this option is switched on then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV