6 transfer, 1 transfer - overview, Transfer – Metrohm tiamo 2.1 Manual User Manual

Page 1153

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1

■■■■■■■■

1139

Event/
Status

Type

Significance

Com-

mands

Start

Event

Command has been started.

All

End

Event

Command has been finished.

All

Dosing

Event

The dosing of the titration has been
started.

DET, MET

Measure

Event

The measurement of the titration
has been started.

DET, MET

Busy

Status

System is in status BUSY, HOLD or
ERROR.

All

Finished

Status

Command has been terminated at
least once.

All

Cond

Status

The command is currently in Con-
ditioning status (COND BUSY or
COND HOLD); the Start drift has
not yet been reached.

SET, KFT,
KFC

Condok

Status

System is in Conditioning status
(COND READY); the Start drift
has been reached.

SET, KFT,
KFC

Note

An event can only be received as such if the RECEIVE command has
been activated at the time of the event. A status, on the other hand, can
be queried at any time.

Comment

Comment on the event message

Input

250 characters

Default value

'empty'

5.6.9.6

TRANSFER

5.6.9.6.1

TRANSFER - Overview

Dialog window: Method

TRANSFER Properties... TRANSFER - 'Command

name'

Command for data transfer via RS-232 to external devices.

Appearance
The command has the following appearance: