Metrohm tiamo 2.1 Manual User Manual
Page 517

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1
■■■■■■■■
503
Range
2 ... 20
Default value
5
Window
A range (window) can be defined on the measured value axis, on the volume
axis or on the time axis. The break-point evaluation will only be carried out
in the defined window. Only the first break-point in the defined window will
be recognized.
Selection
Measured value | Volume | Time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
Window = Time
Range
0 ... 999999 s
Default value
999999 s