Metrohm tiamo 2.1 Manual User Manual

Page 1088

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1074

■■■■■■■■

tiamo 2.1

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

150.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

500.0 °C

Upper hysteresis

Upper hysteresis of the temperature. When the temperature once again falls
below the upper limit by this hysteresis value, the action Temperature
upper limit ok
is triggered

Titrando
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Action

Selection of the event to be triggered if the upper or lower limit is exceeded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is executed.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a mes-
sage appears. As soon as the monitored temperature is again within the
limits (including hysteresis) reagent dosing can be resumed by pressing
on [Continue] in this message box.
Wait for limit ok
Reagent dosing in the current DOS command is interrupted. As soon as
the monitored temperature is once again within the limits (including hys-
teresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.