2 cal loop ph - properties – Metrohm tiamo 2.1 Manual User Manual

Page 1012

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

998

■■■■■■■■

tiamo 2.1

Identifi-
cation

Description

.SLO

Electrode slope calculated from the calibration

5.6.5.4.2

CAL LOOP pH - Properties

Dialog window: Method

▶ CAL LOOP pH ▶ Properties... ▶ CAL LOOP pH - 'Command

name'

The number of buffers and the buffer type are defined in the dialog window
CAL LOOP pH - 'Command name'.

Command name

Name of the command.

Input

25 characters

Buffers

Number of buffers

Selecting the number of buffers for the pH calibration.

Selection

1 | 2 | 3 | 4 | 5

Default value

2

Buffer type

Selecting a predefined buffer series for the pH calibration.

If these buffers are used, tiamo will recognize the particular buffer auto-
matically.

Selection

Metrohm | NIST | DIN | Fisher | Fluka Basel |
Mettler | Merck | Merck CertiPUR 25°C | Beck-
man
| Custom | Special | Radiometer | Baker |
Hamilton DURACAL | Precisa

Default value

Metrohm

Custom
Uses the custom buffer series defined in the configuration under
Tools

▶ Templates ▶ Custom calibration buffers.

Special
Up to five calibration buffers can be defined directly in the Properties
window of the command CAL LOOP pH.