5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
908
■■■■■■■■
tiamo 2.1
Slope
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break-points will be found.
Range
0.0 ... 10.0
Default value
0.9
Smoothing factor
The higher the smoothing factor, the fewer endpoints will be found.
Range
2 ... 20
Default value
5
Window
A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-
dow. Only the first break point in the defined window will be recognized.
Selection
Measured value | time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Time for the lower limit of the window.
Window = Measured value
Range
-200.0 ... 200.0 µA
Default value
-200.0 µA
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Value for the upper limit of the window.
Window = Measured value
Range
-200.0 ... 200.0 µA
Default value
200.0 µA
Window = Time
Range
1 ... 999999 s
Default value
999999 s