5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1022

■■■■■■■■

tiamo 2.1

Max. waiting time

If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:

Range

0 ... 999999 s

Default value

110 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

0.1 ... 999999 s

Default value

2.0 s

Temperature

Temperature

The temperature at which calibration is carried out can be entered manually.
If a temperature sensor is connected and the Temperature measure-
ment
is set on the General/Hardware tab under Sensor to automatic
or continuous, then the temperature will be measured continuously. The
value is used for temperature compensation with the pH measurement.

Range

-20.0 ... 150.0 °C

Default value

25.0 °C

5.6.5.8

CAL Cond

5.6.5.8.1

CAL Cond - Overview

Dialog window: Method

CAL Cond Properties... CAL Cond - 'Command name'

Command for Conductivity standards measurements for the determi-
nation of the cell constants of conductivity sensors.

Devices
This command can be executed with the following devices:

Conductometer: 712, 856

Appearance
The command has the following appearance: