this minimum beaker radius is undercut then the run will be stopped and an
error message will appear. With off no check will be made.

Selection

1.0 … 100.0 mm | off

Default value

off

Lift rate

Lift speed for manual operation.

Range

3 ... 25 mm/s

Default value

25 mm/s

Axial distance

Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.

Range

100.0 ... 300.0 mm

814 USB Sample Processor XL
Selection

166.0 mm

Default value

166.0 mm

815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler, 864 Bal-
ance Sample Processor
Selection

196.0 mm

Default value

196.0 mm

Swing Head
This parameter is only shown when a swing head is mounted on the tower.

Serial number

Shows the serial number of the swing head.

Swing position

Lift position valid for all 4 external positions, at which the swing head turns
to the external positions.

Range

0 ... 235 mm

Default value

0 mm

Rinse position

Rinse position valid for all 4 external positions.

Range

0 ... 235 mm

Default value

0 mm