Metrohm tiamo 2.1 Manual User Manual

Page 830

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

816

■■■■■■■■

tiamo 2.1

-
negative measured value alteration
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Measuring point recording

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Range

0.1 ... 999999 s

Default value

2.0 s

Titrino
Range

1 ... 999999 s

Default value

2 s

Start time

The measured values are not entered in the measuring point list until this
time has elapsed since the start of the titration.

Range

0 ... 999999 s

Default value

0 s

Start measured value pH

The measured values are not entered in the measuring point list until this
value has been reached.

Titrando, 855
Range

-20,000 ... 20,000

Advertising