mxCONTROL Type 8620


Page 66

Calibration Mode “1-point”

This calibration mode can be used for the calibration
of 4-20mA input signals from e.g.

• Redox

Transmitters

• pH-Transmitters

This mode of calibration needs one buffer solution to
receive the new straight calibration line.

Take care for the dependence of temperature
of the buffer solution/pH-measurement.
Usually it is mentioned on the buffer bottles.

Hold the current PV during the calibration procedure: Yes or No

(Default:

Yes).

Hold Function

– for calibration without interrupting the process:

it holds the PV constant during the whole calibration procedure.
Therefor the current PV is used when starting the calibration
procedure with confirming “Next Step”.

→ Proceed

with

“Next Step”,

press key “OK”

→ Then

rinse/clean

the electrode as requested, after that press key “OK”

→ Scroll to “1stPoint”, press key “INPUT”.
→ Put

the

electrode into buffer solution

and input the value of this buffer

solution (care for temperature dependence), confirm with “OK”.

Wait until the measured PV has a steady value, then
→ Proceed

with

“Next Step”,

press key “OK”

The controller measures the current PV of the buffer solution and the corres-
ponding current I_1…

… and calculates Offset and Slope values.
Rinse/clean the electrode again before putting back into process fluid.
→ Finish calibration and the Hold Function of PV with “End”, press key “OK”

Calibration Mode “2-point”

This calibration mode can be used for the
calibration of 4-20mA input signals from e.g.

• pH-Transmitters

This mode of calibration needs two buffer solutions
to receive the new straight calibration line.

Take care for the dependence of tempe-
rature of the buffer solution/ pH-measure-
ment. Usually it is mentioned on the
buffer bottles.