2.1 Setting up the instrument

819 IC Detector / 820 IC Separation Center

13

2 Installation

2.1

Setting up the instrument

2.1.1 Packaging

819 IC Detector and 820 IC Separation Center are supplied together

with the separately packed accessories in special packagings

containing shock-absorbing foam linings designed to provide excellent

protection. The actual instrument is packed in an evacuated

polyethylene bag to prevent the ingress of dust. Please store all these

special packagings as only they assure transport of the instruments

free from damage.

2.1.2 Check

After receipt, immediately check whether the shipment is complete and

has arrived without damage (compare with delivery note and list of

accessories in Section 6.2). In the case of transport damage, see

instructions in Section 6.5.1 "Warranty".

2.1.3 Location

Position the instrument the laboratory at a location convenient for op-

eration, free from vibrations and protected against a corrosive atmos-

phere and contamination by chemicals.

To avoid disturbing temperature influences on the insulated column

compartment, the pump and eluent reservoir must be protected

against direct sunlight.

2.1.4

Arrangement of the instruments

Modular IC instruments can be stacked on top of each other in any se-

quence.

In a modular system any instruments containing liquid-transporting

assemblies should always be located as low down in the stack as

possible, so that any leaks which may occur in tubing or connections

will cause no damage to the other instruments by escaping liquids

such as acids.