6.1 Technical data

819 IC Detector / 820 IC Separation Center

87

6 Appendix

6.1 Technical

data

Provided that nothing to the contrary is mentioned, the

published data are typical values at an ambient tem-

perature of 25°C.

6.1.1

819 IC Detector

Conductivity measurement

Measurement ranges

100, 200, 500, 1000, 2000, 5000, 10'000

µ

S/cm

Full Scale

0.05

10'000

µ

S/cm

Temperature correction

Automatic correction of the measured conductiv-

ity to reference temperature 20°C with the adjust-

able temperature coefficients

(1.5 %/°C or 2.5 %/°C)

Accuracy, absolute value

< 4 % of full-scale value for all ranges under the

following reference conditions:

Adjustment with reference resistance to 1.000 V

(range 1 mS/cm, cell constant 16.7 /cm, temp.

coeff. 2.5 %/°C, frequency 5 kHz, room tempera-

ture) to accuracy <

±

2 mV, measurement accu-

racy < 0.5 % after 30 min warm-up time

Linearity

Up to

±

150 % of each full-scale range

Deviations < 0.5 % of full-scale range

Temperature dependence

Typically 25 ppm/°C

Dependence of cell const..

<

±

1 % over entire adjustment range

Drift (electronic)

< 0.0013 % of selected measurement range

/h/°C

Noise (electronic)

Typically < 0.0003 % of selected meas. range

Damping

2-stage damping (Bessel 4

th

order)

"

Damping off

"

0.25 s (10

90 %)

"

Damping on

"

2.00 s (10

90 %)

Autozero

Function

Automatic zero setting (electronic background

compensation) over entire selectable measure-

ment range

Maximum error

±

0.6 % of full-scale range

Linear range

±

150 % of full-scale range

Advertising