2 Installation

819 IC Detector / 820 IC Separation Center

40

53 Separating column

56 Detector block

57 Inlet capillary for detector block

fixed mounting

58 Mounting rail

59 Column holder (6.2027.0X0)

60 Capillary for syringe 13

PEEK capillary, fixed mounting

61 Sample loop (6.1825.210)

20

µ

L, PEEK, fixed mounting

62 Inlet capillary for injector

PEEK-Capillary 6.1831.010

63 PTFE aspirating tube

fixed mounting

With the one-channel system without suppressor module, the IC sepa-

rating column is installed in the 2.820.0X10 IC Separation Center as fol-

lows (see Fig. 21):

1 Connect column to injector

Remove end caps from column 53.

without column heating:

Screw inlet end of separating column 53 (note flow direction)

to column connection capillary 50 mounted on the injector.

with column heating:

Prepare column heating according to Section 2.9.2 and

screw column connection capillary 50 (see Fig. 20) with a

compression fitting to injection valve 51.

With precolumn:

Install precolumn according to the supplied leaflet between

inlet of the separating column and the injection valve.

2 Rinse column

Place a beaker beneath the column outlet.

Start 818 IC in «IC Net» with suitable flow (see leaflet of the

column) and rinse column for ca. 10 min with eluent.

Stop 818 IC Pump.

3 Connect column to detector block

without column heating

Screw outlet end of separating column 53 to the inlet capillary

57 permanently mounted on the detector block 56.

with column heating

Connect outlet end of separating column 52 (see Fig. 20) with

coupling (6.2744.040) to the inlet capillary 57 permanently

mounted on the detector block 56.

4 Fix column

without column heating

Insert one or two column holders 59 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 58 and fasten separating

column in the column holder.

with column heating

Insert column heating according to Fig. 10 into the Separation

Center.