6 Appendix

819 IC Detector / 820 IC Separation Center

88

Conductivity detector

Construction

Thermostatted conductivity detector with 2 ring-

shaped steel electrodes

Measurement principle

Alternating current measurement with following

cell supply:

Meas. Range

Amplitude

Frequency

100

µ

S/cm

1 V peak

714 Hz

200

µ

S/cm

1 V peak

5000 Hz

500

µ

S/cm

1 V peak

5000 Hz

1 mS/cm

1 V peak

5000 Hz

2 mS/cm

0.5 V peak

5000 Hz

5 mS/cm

0.2 V peak

5000 Hz

10 mS/cm

0.1 V peak

5000 Hz

Effective cell volume

0.8

µ

L

Cell constant

16.7 /cm; adjustable 13.0

21.0 /cm

Maximum back pressure

for measuring cell

5.0 MPa (50 bar)

Thermostatting

Connectable dynamic control to adjustable oper-

ating temperature

Operating temperature

Adjustable in steps of 5°C from 25

45°C

Max. temperature deviation

±

1.5°C

Temperature stability

0.01°C at constant ambient temperature

Parts and controls

Indicator

Power LED 1

ON/Off switch

At rear of instrument

Mains connection

Voltage

115 V: 100...120 V

±

10 %

230 V: 220...240 V

±

10 %

Switching with mains voltage selector in fuse

holder (see Section 2.4.1)

Frequency

50...60 Hz

Power consumption

70 VA

Fuse

5 mm Ø, 20 mm length

100

120 V: 0.63 A (slow-blow)

220

240 V: 0.315 A (slow-blow)

Interfaces

RS 232 interface

Specifications, see Section 5.1

Remote interfaces

Specifications, see Section 5.2

Analog output

Specifications, see Section 5.3