5 Interfaces

819 IC Detector / 820 IC Separation Center

86

5.5 Valve

interfaces

The two valve interfaces 26 and 31 can be used to connect the 820 IC

Separation Center directly to the 830 IC Interface (see Section 2.3.1) or

external devices (e.g. 766 IC Sample Processor or 788 IC Filtra-

tion/Dialysis Sample Processor). The two valve interfaces for the injec-

tion valves A and B (or the suppressor module) are identical and have

the following connections:

820 IC Sep. Center Connection

Function

Inputs

0V

Ground Ground

+5V

10k


Fill

Valve

"FILL"

Valve is switched to the position "FILL"
(or suppressor module is switched further)

H

L

t

P

> 2 ms

+5V

10k


Inject

Valve

"INJECT"

Valve is switched to the position "INJECT"

H

L

t

P

> 2 ms

Outputs

COM


RUN

Position "Fill"
A pulse is outputted when the valve is switched to
the position "FILL".

H

L

t

P

250 ms

COM


RUN

Integrator Start
A pulse is outputted when the valve is switched to
the position "INJECT".

H

L

t

P

250 ms

No liability whatsoever will be accepted for damage arising from the improper connection of devices.