2.9 Separating columns and suppressor module

819 IC Detector / 820 IC Separation Center

41

2.9.6

Two-channel system without suppressor module

Make sure that the connection separating column outlet – detector

inlet is as short as possible in order to prevent unnecessary peak wid-

ening after the separation.

With the two-channel system without suppressor module (2.820.0X20

IC Separation Center), the first IC separating column is connected on

the left side to injection valve A and detector block A as with the one-

channel system (see Section 2.9.5 und Fig. 21). The second column is

connected on the right side to injection valve B and detector block B in

an analogous manner.

Fig. 22:

Interior of the2.820.0320 Separation Center