Fig. 27, Diagram of measuring and full-scale range – Metrohm 820 IC Separation Center User Manual

Page 62

Advertising
background image

3 Operation

819 IC Detector / 820 IC Separation Center

54

Full

scale

The full scale range (operating range) sets the
desired sensitivity for the analog output. The
possible values of the full scale range in

µ

S/cm

depend on the preset measuring

Range

.

This value can be reduced in a total of 11 stages by a factor 1 (low-

est sensitivity) up to maximum a factor of 2000 (highest sensitivity).
The full-scale range is primarily used to increase the sensitivity for

the recording of the chromatogram following the electronic back-

ground compensation of the eluent sensitivity (auto-zero function,

see also description of the auto-zero function above.
For the selected full-scale range (FS), the linearity of the conductivity

measurement is assured in the range –1.8 Full Scale

+1.8 Full

Scale. If the measured value violates these limits, the live display of

Conductivity

in the

WATCH WINDOW

changes it’s color to the settings

for

Out of range

. (default: red). At the same time the absolute conduc-

tivity is displayed instead of the auto-zero value.
The full-scale range determines the limits for the output of the analog

signal at analog output 5 (0

1 V). The polarity of the output signal

can be changed at any time with the parameter

Polarity

to

+

(0

+Full Scale) or

(0

–Full Scale).

Fig. 27:

Diagram of measuring and full-scale range

Full-scale-range

Measuring range

Analog output

Conductivity

0

R

1.1 R

0

+ FS

– FS

+ 1.8 FS

– 1.8 FS

0

+

0 V

10 mV

10 mV

Conductivity Out of Range

abs. Cond.

Out of Range

Conductivity Out of Range