4 Notes - Maintenance - Faults
819 IC Detector / 820 IC Separation Center
78
No feed of regen-
eration or rinsing
solution for sup-
pressor
•
Contamination of the
filter in the 6.2821.120
Filter unit PEEK
•
Counterpressure too
high in suppressor
module
•
Replace the 6.2821.130
filter(see Section 2.7.3)
•
Clean or replace suppres-
sor module (see Section
4.2.6
Detector is not
recognized by «IC
Net»
•
No RS connection
•
Switch instrument on
•
Check wiring
•
Check that the connection
specified in the 819 IC
Detector window (see
Section 3.1.2 Links)
matches the actual wiring