2.2 Installation of accessories

819 IC Detector / 820 IC Separation Center

15

2.2 Installation

of

accessories

2.2.1

Insert detector block

In order to operate an 820 IC Separation Center with an 819 IC Detector

the detector block contained in the standard equipment of the detector

(1.732.0110 Detector block metal-free or 1.732.0100 Metal) must

be inserted in the upper part of the Separation Center.

Fig. 7:

Insert detector block in the Separation Center

2.820.0X10/2.820.0X30 IC Separation Center

The instrument versions 2.820.0X10 and 2.820.0X30 of the IC Separa-

tion Center are operated with one 819 IC Detector. It is best to proceed

as follows when connecting the two instruments and the detector block:

1 Install detector block

Unscrew the four knurled screws 15 from the top rear panel

21 of the 820 IC Separation Center and remove rear panel

(see Fig. 5).

Position detector block from the back in the space provided

in the 820 IC Separation Center on the right and push fully to

the front (see Fig. 7, resp. Fig. 21 and Fig. 23).

Insert the cable permanently attached to the detector block in

opening 20 and the outlet capillary in opening 19 "Waste A"

of the rear panel 21.

Replace rear panel 21 and screw to the 820 IC Separation

Center using the four knurled screws 15.