4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

66

4 Notes - Mainte-

nance - Faults

4.1

Practical notes on ion chromatography

4.1.1 Separating

columns

The attainable quality of analyses with the 819/820 IC system depends

to a large extent on the separation efficiency of the column used. When

purchasing an IC column you should ensure that the separation effi-

ciency suffices for the analysis problems at hand. Ascertain the char-

acteristic data of the IC column on the standard chromatogram en-

closed with the column such as capacity factors, selectivity, plate num-

ber and resolution and check these data with your own measurements.

If any difficulties arise, you should always first check the quality of the

column by recording a standard chromatogram.
Detailed information about the separating columns available from

Metrohm can be found in the data sheet accompanying your separat-

ing column, in the Metrohm IC Column Catalog, which is available

form your local Metrohm agency, or on the Internet under

http://www.metrohm.com

in the ion chromatography product sector. In-

formation about special IC applications can be found in the relevant

"Application Bulletins" or "Application Notes"; these are available

on the Internet under

http://www.metrohm.com

in the applications sec-

tor, or can be requested free of charge from your local Metrohm

agency.

Protection

To protect the column against foreign particles which could have an

adverse influence on the separation efficiency, we advise you to subject

both the eluents and all samples to microfiltration (0.45

µ

m filter) and

to siphon the eluent through the 6.2821.090 Aspirating Filter.
To avoid contamination by abrasive particles arising from piston seals

of the 818 IC Pump, it is advantageous to install an in-line filter be-

tween the pump and the 820 IC Separation Center. The 6.2821.120

Filter unit PEEK 47 (see Section 2.7.3) supplied with the 818 IC Pump

for operation with PEEK capillaries in the pressure range 0

30 MPa.

The use of readily interchangeable precolumns serves to protect the

actual separating columns and increase their service life appreciably.

The precolumn that is suitable for your separating column can be found

in the Metrohm IC Column Catalog which is available from your local

Metrohm agency, the data sheet accompanying your separating col-

umn, the product information about separating columns that can be

found under

http://www.metrohm.com

, ion chromatography products,

or let your agency advise you directly.