2.7 Connection of 818 IC Pump

819 IC Detector / 820 IC Separation Center

31

2.7.4

Connection to injection valve

It is recommended to use 6.1831.010 PEEK capillaries, a 6.2620.150

Pulsation dampener (see Section 2.7.2) and a 6.2821.120 Filter unit

PEEK (see Section 2.7.3) to connect the 818 IC Pump to the injection

valve of the 820 IC Separation Center. Proceed as follows:

Fig. 19:

Connection to injection valve

3 Connection capillary

to 818 IC Pump

PEEK capillary (6.1831.010)

47 Filter unit PEEK (6.2821.120)

48 PEEK capillary (6.1831.010)

49 Pulsation dampener 6.2620.150

50 Column connection capillary

PEEK capillary (6.1831.010)

51 Injection valve

1 Connection to 818 IC Pump

Cut connection capillary 3 (6.1831.010 PEEK capillary) to the

required length and equip it with compression fittings.

Attach connection capillary 3 to the upper connection of the

purge valve (see Fig. 1 818 Instructions for Use).

2 Connection of filter unit PEEK

Attach the other end of connection capillary 3 at the filter

housing 44 of the filter unit 47 (see Fig. 18).

Connect a PEEK capillary 6.1831.010 cut to the required

length and equipped with connectors to the filter connector

46 of the filter unit 47.

49

47

3

48

50 51

48

818

820