5 interfaces – Metrohm 820 IC Separation Center User Manual

Page 92

Advertising
background image

5 Interfaces

819 IC Detector / 820 IC Separation Center

84

819 IC Detector

Pin

Function

Outputs

5 Valve

A

"FILL"

Valve A is switched to position "FILL".

H

L

t

P

> 50 ms

18 Valve

A

"INJECT"

Valve A is switched to position "INJECT".

H

L

t

P

> 50 ms

4

Lock Terminal block 26, 31
L = functions <FILL> and <INJECT> are locked
H = functions <FILL> and <INJECT> are not locked


V

CEO

= 40 V

IC =

20

mA

17 Valve

B

"FILL" or suppressor

"STEP"

Valve B is switched to position "FILL" or suppressor
module is switched to next position.

H

L

t

P

> 50 ms

3 Valve

B

"INJECT"

Valve B is switched to position "INJECT".

H

L

t

P

> 50 ms

2 Remote

6

Programmable remote output line 6
(command "remote" in program or event)

V

CEO

= 40 V

IC =

20

mA

1 Remote

7

Programmable remote output line 7
(command "remote" in program or event)

V

CEO

= 40 V

IC =

20

mA

16 Remote

8

Programmable remote output line 8
(command "remote" in program or event)


V

CEO

= 40 V

IC =

20

mA

Voltages

+5V

0V

24VAKT

0VAKT

15
14
25
6
13
8
19
20


5 V: inactive

High (H)

0 V: active

Low (L)


I

40

mA

R

i

12

Contact arrangement at socket
"IC Separation Center" (female)

13

1

25 14


Connecting cables to external devices must have an appropriate
25-pin connector (male). Metrohm offers connector sockets
(K.210.9060) and the associated housing (K.210.0002) as an
option for the preparation of such cables.

No liability whatsoever will be accepted for damage arising from the improper connection of devices.