4.2 Maintenance and servicing

819 IC Detector / 820 IC Separation Center

69

Eluent change

When the eluent is changed, it must be ensured that no precipitates

can be formed. Solutions used in direct succession must therefore be

miscible. If the system has to be rinsed with an organic solution, several

solvents with increasing or decreasing lipophilic character may possibly

have to be used (e.g. water

acetone

chloroform).

4.1.4 Suppressor

module

To avoid contamination of the suppressor module by foreign particles

or bacterial growth, it is advantageous to install an in-line filter be-

tween the 833 IC Liquid Handling Pump Unit and the suppressor mod-

ule. The two 6.2821.120 Filter units PEEK (see Section 2.7.3 und

Section 2.9.4) supplied with the 2.833.0010 IC Liquid Handling Pump

Unit are eminently suitable for this purpose.

4.1.5 Connections

All connections between injector, column and detector must be as short

as possible, have a low dead volume and be absolutely tight. The PEEK

capillary after the detector block must be free from constriction (the

measuring cell is tested to 5 MPa = 50 bar back pressure).

4.2 Maintenance

and

servicing

4.2.1 General

information

Care

The 819 IC Detector and 820 IC Separation Center require proper care

and attention. Excessive contamination of the instruments could possi-

bly lead to malfunctions and a shorter service life of the inherently rug-

ged mechanical and electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Although constructional measures have been designed to virtually

eliminate such a situation, should corrosive media penetrate the inte-

rior of the instruments the mains plug of the 819 IC Detector must be

immediately disconnected to prevent extensive damage to the instru-

ment electronics. Inform Metrohm service if your instrument(s) have

been damaged in such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in Section 1.4.1.