2.9 Separating columns and suppressor module

819 IC Detector / 820 IC Separation Center

49

2 Check 818 IC Pump

Check that the aspiration capillary of 818 IC Pump is im-

mersed in the eluent.

In the loaded system open the

818 IC Pump

window and

check/set the values for the flow rate and the maximum

switch-off pressure under

Settings

.

The recommended flow rate for the separating column used

can be found in the corresponding column data sheet (nor-

mally 0.5

2 mL/min).

The maximum switch-off pressure should normally be approx.

3 MPa above the pressure observed with the used column,

for details see 818 Instructions for Use.

Start 818 IC Pump with

<Start>

.

3 Check for leaks

Check all capillaries and their connections between the 818

IC Pump and the detector block for escaping liquid. If eluent

escapes anywhere, the appropriate compression fitting must

be tightened further or changed..

Check also all tubing’s and their connections between the

833 IC Liquid Handling Unit and the suppressor module (if

present) for escaping liquid and tighten or change the fittings

if necessary.

4 Check 819 IC Detector

Open the window

819 IC Detector

of the loaded system and

check/set the values of operating temperature, cell constant,

measuring range and full-scale range on the

Method Parameter

tab (see sec. 3.1.2).

Set measuring range

Range

so that the displayed absolute

conductivity value of the eluent lies within the selected range

(default value = 1 mS/cm).

Set full-scale range,

Full

scale

(default value = range/1). To

start with, it is advisable not to set too narrow a full-scale

range as the conductivity of the eluent in the conditioning

phase can change greatly until attainment of a stable tem-

perature.

5 Condition IC system

Rinse the IC system with eluent until the desired stability of

the baseline is reached (normally 30

60 min).

6 Condition suppressor (if present)

Open window

820 IC Separation Center

and switch suppressor

module at least three times to the next position every 10 min.