Fig. 24, Suppressor module connections, Fig. 24 ) – Metrohm 820 IC Separation Center User Manual

Page 52

Advertising
background image

2 Installation

819 IC Detector / 820 IC Separation Center

44

3 Connect column to suppressor module (Suppressor

connection 1)

Cut inlet capillary 64 (marked with "Eluent") of suppressor

module 70 to the required length. Use the 6.2126.080 Capil-

lary cutter available as an option.

without column heating

Screw inlet capillary 64 on to the outlet end of separating col-

umn 53 using a 6.2744.010 Compression fitting.

with column heating

Connect capillary 52 at the outlet of the separating column

(see Fig. 20) to the inlet capillary 64 using the 6.2744.040

Coupling.

4 Fix column

without column heating

Insert one or two column holders 59 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 58 and fasten separating

column in the column holder.

with column heating

Insert column heating according to Fig. 10 into the Separation

Center.

Fig. 24:

Suppressor module connections


64 Suppressor inlet capillary for

eluent

65 Suppressor inlet capillary for

H

2

SO

4

66 Suppressor outlet capillary

for H

2

SO

4

67 Suppressor outlet capillary

for H

2

O

68 Suppressor inlet capillary for

H

2

O

69 Suppressor outlet capillary

for eluent

Eluent

H

2

O

Detector

Waste

Waste

H

2

SO

4

1

3

2

65

64

69

66

68

67