3 analog output, Analog output, External power supply for 820 ic separation center – Metrohm 820 IC Separation Center User Manual

Page 93

Advertising
background image

5.3 Analog output

819 IC Detector / 820 IC Separation Center

85

5.3 Analog

output

Analog output 5 "Output“ (0

1 V) of the 819 IC Detector (see Fig. 3) is

used for connecting a recording channel to 830 IC Interface or 771 IC

Compact Interface or from an external recorder or integration system.

The connection from 830 IC Interface and 771 IC Compact Interface is

described in Section 2.3. The polarity of the output signal at the analog

output connectors can be changed at any time under

Polarity

in the

window

819 IC Detector

in «IC Net» (see Section 3.1.2). The switching of

the analog output is as follows:

819 IC Detector

Pin

Voltage

Contact arrangement at
socket "Output" (female)

5

1

9

6


1

6


0 V

Signal 0...1 V

The other pins are not assigned.

5.4

External power supply for
820 IC Separation Center

If the 820 IC Separation Center is operated without the 819 IC Detector,

an external power supply must be attached to connection 25 "External

Power". Power supply and connecting cable must comply with the fol-

lowing conditions:

"External Power" socket

Function

1

4

5

3

6

7

2

1

+5 V DC / 0.5 A

2 not

assigned

3

+24 V DC / 2 A

4

0 V (digital) *

5

0 V (analog) *

6 Ground
7 Ground

* Pin 4 and pin 5 must

be connected in the
vicinity of the power
source ("zero point of
star")