Fig. 29, Assembling the suppressor, Fig. 29 ) – Metrohm 820 IC Separation Center User Manual

Page 82: 72 from suppressor holder 75, Pull out connection piece 73 and suppress, R 74 from, R holder 75 (the connection, Ection piece 73 from suppress, R 74, N piece 73

Advertising
background image

4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

74

6 Clean connection piece

Clean the sealing surface of connection piece 73 using a lint-

free cloth and ethanol.


Fig. 29:

Assembling the suppressor

72 Screw nut

73 Connection piece with input and

output leads

(6.2832.010)

for suppressor module

74 Suppressor rotor(6.2832.000)

75 Suppressor holder

7 Insert connection piece

Insert connection piece 73 in suppressor holder 75 in such a

way that connection "1" is at the top, and that the three lugs

on the connection piece fit in the corresponding openings of

the holder.

Screw nut 72 onto the thread of suppressor holder 75 manu-

ally (do not use tools).

8 Connect and condition the suppressor

Reconnect the suppressor to the IC system.

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.

74

73

72 75

Advertising
This manual is related to the following products: