8 replacement of suppressor module, Replacement of suppressor module, Ction 4.2.8 ) – Metrohm 820 IC Separation Center User Manual

Page 83

Advertising
background image

4.2 Maintenance and servicing

819 IC Detector / 820 IC Separation Center

75

4.2.8

Replacement of suppressor module

The suppressor in the suppressor block may have to be replaced in the

following cases:

Irremediable loss of suppressor capacity (reduced

phosphate sensitivity and/or strong rise in baseline)

Irremediable blockage of the suppressor (the suppres-

sor can no longer deliver solutions)

Both the 6.2832.000 Suppressor rotor and the 6.2832.010 Connection

piece with the input and output leads can be replaced. To replace these

components proceed as follows (see Fig. 29):

1 Disconnect suppressor from IC system

Disconnect all input and output leads of the suppressor from

IC system and the 833 IC Liquid Handling Pump Unit.

2 Dismantle suppressor

Unscrew nut 72 from suppressor holder 75.

Pull out connection piece 73 and suppressor rotor 74 from

suppressor holder 75 (the connection piece and the rotor

normally stick together).

Loosen connection piece 73 from suppressor rotor 74.

3 Clean suppressor rotor

Clean the sealing surface of new suppressor rotor 74

(6.2832.000) using a lint-free cloth and ethanol.

4 Insert suppressor rotor

Insert suppressor rotor 74 in suppressor holder 75 in such a

way that the tubing connections at the rear of the rotor fit in

the corresponding openings inside the rotor, and that one of

the three holes in the rotor can be seen from below in one of

the openings of the holder.

If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

5 Clean connection piece

Clean the sealing surface of connection piece 73 using a lint-

free cloth and ethanol.

Advertising
This manual is related to the following products: