3 Operation

819 IC Detector / 820 IC Separation Center

52

Zero <ON>

Triggers the auto-zero function.

Zero <OFF>

Switch off the auto-zero function.

"Auto-zero" is the name used for the automatic electronic back-

ground compensation, i.e. the current measured

Conductivity

value

is set to zero and hence lies in the middle of the selected full-scale

range.
The auto-zero function is disabled, if the actual value of the absolute

conductivity

abs. Cond.

out of the limit set by the measuring

Range

(see Fig. 27), the diagram below shows the operating principle of the

auto-zero function.

Fig. 26:

Operating principle of Auto-zero

Before the auto-zero function is activated for the first time or after it

has been switched off, the measuring signal lies outside the full-

scale range; it is automatically set to 0

µ

S/cm each time the

Zero

<ON>

key is pressed.

The auto-zero function can be triggered at any time using the

Zero

<ON>

key.

The auto-zero function works properly only when the measured

value remains relatively stable (e.g. not during the appearance of

the injection peak at the start of the chromatogram).

Zeit

0

+

Full Scale

Full Scale

+ 1.8 Full Scale

– 1.8

Full Scale

Conductivity

Conductivity
Out of range

Conductivity
Out of range

Conductivity
Out of range

Zero <OFF>

Zero <ON>

Zero <ON>