3.1 819 IC Detector

819 IC Detector / 820 IC Separation Center

57

Program

On the

Program

subpage, program steps including time, program in-

struction and parameter can be entered.

First column

Time at which program instruction is applied.

Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is
applied together with the last instruction with
time entry.

Second column

Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the

Remote configuration

tab.

Third column

Parameter for program instruction (see below).

ENABLED

Enable program start (a disabled program is

not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are dis-
played if program is wrong).


List of program instructions

The following program instructions can be added to the time program
on the

Program

subpage:

Instruction Parameter entry

Meaning

Remote

0, 1, *

Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter

1

,

0

or

*

. For entry of the

other values, move the cursor in front
of the value to be changed and enter

1

,

0

or

*

.

Range

100, 200, 5000

µ

S/cm

1, 2, 5, 10 mS/cm

Set measuring range to the selected
value.

FS 0.05

µ

S/cm ...

10 mS/cm

Set full scale range to the selected
value (depends on Range setting).

Zero on,

off

Switch on or off the auto-zero func-
tion.

Polarity +,

Reset the polarity of the output sig-
nal.

Mark

Trigger marking signal.