2 passivation, 3 recycling, 4 shutdown – Metrohm 820 IC Separation Center User Manual

Page 78: Passivation, Recycling, Shutdown

Advertising
background image

4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

70

Maintenance by Metrohm service

Maintenance of the 819 IC Detector and 820 IC Separation Center is

best done as part of an annual service performed by specialists from

the Metrohm company. If work is frequently performed with caustic and

corrosive chemicals, it may be necessary to shorten the interval be-

tween servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

4.2.2 Passivation

Passivation of the entire IC system (without column) by rinsing with

20...50 mL 0.2 mol/L HNO

3

is appropriate only in the following case:

if exceptional changes are observed in the

measurement properties of the cell

For passivation the separating column is removed from the 820 IC

Separation Center. The two capillaries 50 and 57 (see Fig. 21 and Fig.

23) are connected directly to each other with the 6.2744.040 Coupling

71 included in the accessories.

4.2.3 Recycling

To keep the eluent consumption between injections to a minimum when

the system is at rest (e.g. overnight), the so-called recycling procedure

can be used. In recycling the eluent exiting the outlet capillary of the

detector block is led back directly to the eluent supply vessel. The IC

system is thus quickly ready for new injections without a long condition-

ing period.

The recycling procedure must

not be used

in operation with the suppressor module,

with alkaline eluents,

with the 6.1010.000 IC Cation column METROSEP Cation 1-2.

4.2.4 Shutdown

If the 819/820 IC-System is shut down for a considerable length of time,

the entire IC system (without column and suppressor) must be rinsed

free from salt with methanol/water (1:4) to avoid crystallization of elu-

ent salts with the corresponding subsequent damage.
For rinsing the separating column is removed; the two capillaries 50

and 57 (see Fig. 21 and. Fig. 23) are directly connected to each other

with the 6.2744.040 Coupling enclosed with the accessories. Rinse with

methanol/water (1:4) until the conductivity drops below 10

µ

S/cm.