2 Installation

819 IC Detector / 820 IC Separation Center

38

To avoid contamination of the suppressor module by foreign particles

or bacterial growth, it is advantageous to install an in-line filter be-

tween the 833 IC Liquid Handling Pump Unit and the suppressor mod-

ule. For this the two 6.2821.120 Filter units PEEK (mounting see Fig.

18 Section 2.7.3) supplied with the 2.833.0010 IC Liquid Handling

Pump Unit should be used.

The suppressor module must never be switched in the dry state as

there is a danger of blocking. Before every switching operation of the

suppressor module, the three suppressor units must have been

rinsed for at least ½ h with eluent, regeneration and rinsing solution.