Fig. 28, Flow path of the injection valve, See fig. 28 ) of – Metrohm 820 IC Separation Center User Manual

Page 70: Fig. 28 ) of

Advertising
background image

3 Operation

819 IC Detector / 820 IC Separation Center

62

Auto Step

*

fill, inject, off

Display of current auto step status.

Entry field

Set automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection:

fill, inject, off

* only available for

820 IC Separation Center

connected to 819

Fig. 28:

Flow path of the injection valve

If the injection valve is in the

<FILL>

position, the sample loop 61 is

filled by siphoning in the sample solution using the syringe. The eluent

flows directly to the separating column.
If the injection valve is in the

<INJECT>

position the eluent flows via the

sample loop to the separating column. This causes injection of the

sample previously added to the sample loop. Sample inlet tubing and

syringe are directly connected.

1

2

4

5

3

6

Syringe

Column

Eluent

Sample

Loop

1

2

4

5

3

6

Syringe

Column

Eluent

Sample

Loop

<FILL>

<INJECT>