6 capillary connections, Capillary connections, Fig. 16 – Metrohm 820 IC Separation Center User Manual

Page 36: Connect capillaries

Advertising
background image

2 Installation

819 IC Detector / 820 IC Separation Center

28

2.6 Capillary

connections

Some of the connections under high pressure between the feed pump

and the detector block must be set up by the user. For that the

6.1831.010 PEEK capillary

(i.d. = 0.25 mm,

o.d. =

1

/

16

"

,

length = 3 m) and the 6.2744.010 Compression fittings supplied in

the accessories of the 820 IC Separation Center can be used. These

capillaries can be used in the pressure range from 0

30 MPa

(0

300 bar).

Capillaries fitted with new connectors must have a perfectly flat cut

surface. For PEEK capillaries it is best to use the

6.2621.080 Capil-

lary tubing cutter

Fig. 16:

Connect capillaries

42 PEEK capillary6.1831.010

43 Compression fitting 6.2744.010


For the connection of 6.1831.010 PEEK capillaries or of the 6.1834.000

PTFE aspirating tubing, the supplied 6.2744.010 PEEK Compression

fittings are used. It is best to proceed as follows:

1 Mount compression fitting

Slide a compression fitting 43 (6.2744.010) over the end of the

capillary 42 to be fastened as shown in Fig. 16.

2 Insert capillary in connection

Push capillary end in the corresponding connection as far as it

will go (to avoid dead volume).

3 Tighten compression fitting

Tighten compression fitting 43 by hand (never use tools).

43

42