3.1 819 IC Detector

819 IC Detector / 820 IC Separation Center

53

If the measured conductivity value exceeds

±

180% of the set full-

scale range, the live display of

Conductivity

in the

WATCH WINDOW

changes it’s color to the settings for

Out of range

. (default: red). At the

same time the absolute conductivity is displayed instead of the auto-

zero value.

Method parameter

The

Method parameter

tab of the

819 IC Detector

window contains the

method startup values for the instrument. These values are sent and

applied to the 819 IC Detector by

Start determination

and

Startup hard-

ware

, or if the values are sent manually with

<SET>

.

Range

Measuring range in

µ

S/cm

Selection:

100, 200, 500

µ

S/cm

and

1, 2, 5, 10

mS/cm

Select the measuring range so that the conduc-

tivity value of the eluent used is always within
the selected range. If the measured conductiv-
ity value exceeds the upper range limit by more
than 10%, the live display of

abs. Cond.

in the

WATCH WINDOW

changes it’s color to the set-

tings for

Out of range

. (default: red). The se-

lected measuring range limits the possible set-
tings of the full-scale range, which can be se-
lected to be maximum 2000 times more sensi-

tive than the measuring range.