5.1 RS 232 interface

819 IC Detector / 820 IC Separation Center

79

5 Interfaces

5.1

RS 232 interface

The RS 232 interface 9 of the 819 IC Detector is used for the connection

of the 830 IC Interface or a PC (see Section 2.3.1 – 2.3.2) and allows

remote operation of 819 IC Detector and 820 IC Separation Center.

5.1.1

Data transmission protocol

The RS 232 interface 9 of the 819 IC Detector is configured as DTE

(Data Terminal Equipment) with the following technical specifications:

Standard

Data interface in accordance with EIA standard

RS 232C (DIN 66020, page 1)

Baud rate 9600

Data Bits

8

Stop Bits

1

XON/XOFF

on

Control characters C

R

DEC 13

HEX 0D

L

F

DEC 10

HEX 0A

XON

DEC 17

HEX 11

XOFF

DEC 19

HEX 13

Max. line length

80 characters + C

R

L

F

Mode

Full duplex (simultaneous transmission and re-

ceiving)
Restriction:

If data are first received by the interface, trans-

mission is not started until the receipt is at an

end.

Cable

For interconnection of the 819 IC Detector with

third-party devices, only a shielded data cable

may be used (e.g. Metrohm D.104.0201). The

cable shield must be perfectly earthed at both

devices (pay attention to current loops: always

use star earthing). Use only connectors with

sufficient shielding (e.g. Metrohm K.210.0001

with K.210.9004).

Cable length

max. ca. 15 m