5.2 Remote interfaces

819 IC Detector / 820 IC Separation Center

83

5.2.2

"IC Separation Center" interface

The 820 IC Separation Center is normally attached to remote interface

10 "IC Separation Center" (see Section 2.3). If the 819 IC Detector is

operated without a Separation Center, this remote interface can also be

used for the connection of external devices. The remote interface "IC

Separation Center" has the following pin assignment:

819 IC Detector

Pin

Function

Inputs

+5V

21

Position "FILL" at injection valve A
L = Valve A in position "FILL"
H = Valve A not in position "FILL"

+5V

9

Position "INJECT" at injection valve A
L = Valve A in position "INJECT"
H = Valve A not in position "INJECT"

+5V

22

Position "FILL" at injection valve B or "in position" at suppressor
L = Valve B in position "FILL" or suppressor set to "in position"
H = Valve B not in position "FILL" or suppressor not set to
"in

position"

+5V

10

Position "INJECT" at injection valve B
L = Valve B in position "INJEC

T"

H = Valve B not in position "INJECT"

+5V

+5V

23

11

2

0

2

1

Configuration of the 820 IC Separation Center

Input

Decimal Status

11 23

code

H

H

0

undefined

H

L

1

valve

A

L

H

2

valve A + suppressor

L

L

3

valve A + valve B

+5V

24

Connection 820 IC Separation Center
L = 820 IC Separation Center connected
H = 820 IC Separation Center not accessible

+5V

12 not

assigned