background image

2 Installation

819 IC Detector / 820 IC Separation Center

30

nectors supplied or two 6.2744.010 PEEK compression fittings can be

used. The flow direction is arbitrary. The pulsation dampener is posi-

tioned in the interior of the 820 IC Separation Center on the base below

the injection valve (see Fig. 21, Fig. 22, Fig. 23, resp. Fig. 25).

The pulsation dampener is filled with isopropanol when new. Rinse

your IC system carefully after the first installation of a new pulsation

dampener.

2.7.3

Filter unit PEEK

The 6.2821.120 Filter unit PEEK 47 supplied (see Fig. 18) serves to

avoid contamination of the piston seals by abrasive particles and can

be used in the pressure range 0

30 MPa (0

300 bar). The filter unit

consists of the housing 44, the filter 45 and the connector 46 to be

screwed into the housing 44. For the connection of capillaries 42, PEEK

compression fittings 43 (6.2744.010) must be used.
New filter 45 are available as an option with the ordering number

6.2821.110 (10 pieces).

For the connection of the filter unit, please note the flow direction ar-

row printed on the housing.
The filter unit is filled with isopropanol when new. Rinse your IC system

carefully after the first installation of a new filter unit.

Fig. 18:

Filter unit PEEK (6.2821.120)

42 PEEK capillary (6.1831.010)

43 Compression fitting (6.2744.010)

44 Filter housing

45 Filter

6.2821.130

46 Filter connector

43

42 43

46

45

44 42

Advertising