3.1 819 IC Detector

819 IC Detector / 820 IC Separation Center

51

Conductivity

Live display of measured conductivity (auto-
zero value) in

µ

S/cm. This field is available for

the

WATCH WINDOW

. The watch window color is

changed to the

Out of range

color if the meas-

ured value is more than 80% outside the

Full

Scale

range

.

abs. Cond.

Live display of absolute conductivity value in

µ

S/cm. This field is available for the

WATCH

WINDOW

. The watch window color is changed

to the

Out of range

color if the measured value is

bigger or equal to 110% of

Range.

Ion

Display of ion settings:

Anion

:

polarity = +, temperature coeff. = 2.5

Cation

:

polarity = –, temperature coeff. = 1.5

Range

Display of measuring range in

µ

S/cm.

Full

scale

Display of full scale range in

µ

S/cm.

Cell

const

. Display

of cell constant of the conductivity cell

in /cm.

Thermostat

Display of temperature within in the detector

block.

Offset

Display of offset of the zero point.

Damping

Display of electronic damping of the analog
output.

Remote lines

Display value of remote line settings 1

8.