KEYENCE IV Series User Manual

Page 102

Advertising
background image

4-34

- IV Series User's Manual (Monitor) -

12

Tap the [Limit Adjustment] button.

The sensor turns to the Test mode.

13

Image a high-quality-target and a low-

quality-target.

If the target is to be imaged using an external
trigger, input the external trigger.
If an external trigger cannot be input, tap the

button (Page 4-7) to image the target by

inputting an internal trigger temporarily. Imaging
by inputting an internal trigger finishes by the

button.

14

Adjust the threshold of anomaly detection

by checking the matching rate.

<Setting example>
If the matching rate of the high-quality-target is

higher than 90 and if the matching rate of the low-

quality-target is lower than 40, set the threshold to 65

which is an intermediate value between 40 and 90.

Matching rate

Threshold

y

A target is detected as OK if the matching

rate is greater than the threshold, and

detected as NG if the matching rate is lower

than the threshold.

y

According to the status result of the tool, the

display color of the tool changes.

“Selecting a display method for tools” (Page 5-7)

y

The threshold can be changed using the

slider or by inputting values.

“Editing the value with the slider” (Page 3-14)

y

The thresholds can also be set in [Limit

Adjustment] of the Sensor Setup Menu screen.

“Adjusting Thresholds for Judgment” (Page 5-14)

y

For details of the matching rate, refer to

“Matching Rate of the Outline Tool and

Position Adjustment Tool” (Page A-5).

15

After the setting is completed, tap the [OK]

button.

The system returns to the main screen for the
Outline tool settings.

16

Tap the [OK] button.

The system returns to the main screen for the

Tool settings.

Outline tool

Color Area/Area tool

Position Adjustment tool

Extended functions

1. Image Optimization

2. Master Registration

3. Tool Settings

4. Output Assignment

4

Settings Navigator (Setting the Judgment Condition)

Advertising