To achieve stated results, the following thermocouples must be used with the INPUT TYPE/RANGE set to TC NARROW

4

REF+LIN

+CAL

°

C

0.55
1.12
2.98

1.45
1.90

1.87

2.39

TC

TYPE

T

R

S

B

See Note:

RANGE

°

C

0/400

-200/0

-250/-220

800/1700

200/800

250/1750

200/1800

1

REF

ACC'Y

°

C

0.12
0.23
0.73

0.43
0.58

0.56

0.74

2

LIN.

ACC'Y

°

C

0.09
0.19
0.36

0.19
0.25

0.25

0.31

3

FACTORY

CAL

°

C

0.34
0.70
1.90

0.83
1.08

1.05

1.34

5

DEVIATION

ACC'Y

°

C

0.04
0.02
0.00

0.09
0.06

0.15

0.16

6

RESOL

°

C/bit

0.018
0.036
0.111

0.065
0.088

0.086

0.113

ACCURACY TABLES CONT.

4

REF+LIN

+CAL

°

C

0.33
0.46

0.32

0.66

0.21

RTD

TYPE

385
DIN

392

USA

392

SAMA

100 ohm

Nickel

See Note:

RANGE

°

C

-160/480

-200/-160

-100/450

-200/560

-40/200

1

REF

ACC'Y

°

C

0.16
0.14

0.16

0.29

0.09

2

LIN.

ACC'Y

°

C

0.03
0.20

0.03

0.13

0.05

3

FACTORY

CAL

°

C

0.13
0.12

0.13

0.24

0.07

5

DEVIATION

ACC'Y

°

C

0.06
0.00

0.05

0.06

0.02

6

RESOL

°

C/bit

0.025
0.022

0.025

0.044

0.013

Advertising