West Control Solutions MRC 9000 User Manual

Page 229

Advertising
background image

Edition 3

Rev G

Section 11

11-7

CHART

Will place a dot on the inner or outer ring, whichever is selected for calibration

SCROLL

Will select either the inner or outer ring

UP

Will move the selected (inner or outer) ring position to the left and print a dot in the new

position without moving the chart.

DOWN

Will move the selected (inner or outer) ring position to the right and print a dot in the new

position without moving the chart.

RESET

Will move the chart 30 steps while not printing any dots

ESC

Will exit with no change to the calibration values

ENTER

Will exit with the new values being stored

Step 3

With INNER RING CALIBRATE on the upper display line, depress the

to view the current location of the

inner ring zero point. Use UP/DOWN to position the dot accordingly, use RESET to move the chart.

Step 4

D e p r e s s

the to toggle the upper display line to read OUTER RING CALIBRATE.

Step 5

D e p r e s s

the to view the current location of the outer ring zero point. Use UP/DOWN to position the dot

accordingly, use RESET to move the chart.

Step 6

D e p r e s s

the to save calibration values.

11.5 MILLIAMP OUTPUT CALIBRATION

This routine is used to adjust the mA Output(s) for span. A milliamp meter is required to perform this calibration.

Step 1

With mA OUTPUTS in the lower display line, depress the

and CAL WHICH mA OUTPUT

appears in the upper display line and 1 appears in the lower display line. To

change,

depress the , then the to desired output number, then depress .