Section 11

When installing an Input Board for any reason, check the location (where this board is mounted), then verify the

position of JU13 and/or JU14. TC WIDE or TC NARROW were selected, go to Step 5, if not, go to Step 6.

Step 5

Depress the

and the upper display line will show IVx CAL CJC OR TC and the lower display line will show

TC, or CJC. Select TC by depressing the ,

then the

until TC is blinking, then depress the

.

Step 6

Depress the

and the upper display line will show IVx PERFORM CAL and the lower display line will show

NO. Select YES to perform the calibration. Connect the appropriate one of the following inputs to the input

terminals of the input being calibrated:

100 ohm resistor

RTD Inputs

0.0 mA DC

Current Inputs

0.0 mV DC

TC or mV DC Inputs

0.0 V DC

Volt Inputs

Wait one minute, then depress the

key.

Step 7

Connect the appropriate one of the following inputs to the input terminals of the input being calibrated:

277 ohm resistor

RTD Inputs

20.00 mA DC

Current Inputs

25.00 mV DC

TC or 0/25 mV DC Inputs

100.00 mV DC

TC or 0/25+ mV DC Inputs

1.00 V DC

1 V DC Inputs

10.00 V DC

10 V DC Inputs

Wait one minute, then depress the

key.

1, 3, 5, 7

JU13

2, 4, 6, 8

JU14

FOR INPUTS
BOARD POSITION

1

2

3

4

Board 4

Board 3

Board 2

Board 1

SIDE
VIEW

Advertising