Edition 3
Rev G

11-6

Section 11

Step 7

Connect the proper TC type up to the input being calibrated. Place the thermometer next to the HOT Junction

(sensor point) of the TC connected. Depress the

and the upper display line will display IVx CAL CJC and

lower display line will display CJC READS: 25.0 C. Depress the

, then use the UP/DOWN to change the

display to read what the mercury thermometer is reading, then depress the

.

Step 8

Depress the

and the upper display line will display IVx PERFORM CAL. Depress the

and select

YES to perform calibration of the CJC.

Wait one minute, then depress the

key.

Step 9

Repeat steps 1 through 8 for each input board to be calibrated.

11.3 PARAMETER DEFAULTS

This routine will invoke all of the default program parameter values. See Appendix C for the listing of default

values.

Step 1

With PARAMETER DEFAULTS in the lower display line, depress the

and SET DEFAULTS will appear in

the upper display line and NO in the lower display line.

Step 2

Depress then select YES to set the defaults.

11.4 CHART CALIBRATION

Step 1

With CHART CALIBRATION in the lower display line, depress

to enter the Chart Calibration. The upper

display line will display START CALIBRATION. To perform the chart calibration, depress

and select YES,

then depress the

. Press any key to continue.

Step 2

The upper display line will display INNER RING CALIBRATE and the lower display line will display SCROLL

TOGGLES RING. The keys will function as follows: