AC Flatness Adjustment Procedure (cont’d)

Example Program (cont’d)

3080 FOR I=1 TO Max_con
3090 Cal_real(I)=Cal_reflect(I)
3100 IF I=1 THEN Cal_real(I)=Cal_reflect(I)/Scale(1)

! M_plus

3110 IF I=2 THEN Cal_real(I)=Cal_real(I)/Scale(11)

! P_base

3120 IF I=3 THEN Cal_real(I)=Cal_reflect(I)/Scale(2)

! M_minus

3130 IF I=4 THEN Cal_real(I)=Cal_reflect(I)/Scale(3)

! M_adj

3140 IF I=5 THEN Cal_real(I)=Cal_real(I)/Scale(11)

! N_base

3150 IF I>5 AND I<13 THEN Cal_real(I)=Cal_real(I)/Scale(10) !Filter and
3160

! ATTN gain errors

3170 IF I=13 THEN Cal_real(I)=Cal_reflect(I)/Scale(4)

! M_off

3180 IF I=14 THEN Cal_real(I)=Cal_reflect(I)/Scale(5)

! M_cust

3190 IF I=15 THEN Cal_real(I)=Cal_reflect(I)/Scale(6)

! Vpwro

3200 IF I=16 THEN Cal_real(I)=Cal_reflect(I)/Scale(6)

! Vpwri

3210 IF I=17 THEN Cal_real(I)=Cal_reflect(I)/Scale(7)

! Vbuf

3220 IF I=18 THEN Cal_real(I)=Cal_reflect(I)/Scale(8)

! M_sum

3230 IF I=19 THEN Cal_real(I)=Cal_reflect(I)/Scale(9)

! B_sum

3240

! Z_inc_0db (not scaled)

3250

! Z_inc_14db (not scaled)

3260 IF I>21 THEN Cal_real(I)=Cal_real(I)/Scale(10)

! Zout gain errors

3270 NEXT I
3280 !
3290 DEALLOCATE Scale(*),Cal_reflect(*)
3300 SUBEND
3310 !
3320 Valid_cons:SUB Valid_cons(Results(*),INTEGER Cal_problem)
3330 COM @Afg,@Dmm,@Pwr_mtr,@Analyzer,Secure_code$
3340 COM /Flat/ INTEGER Num_points,Max_con
3360 Max_con=25
3370 Cal_problem=0
3380 !
3390 ALLOCATE Cal_real(1:Max_con)
3400 !
3410 CALL Read_dc_cal_con(Cal_real(*))
3420 !CHECK FOR VALID CAL
3430 M_plus=Cal_real(1)

! key cal constant

3440 P_base=Cal_real(2)

! +base

3450 M_minus=Cal_real(3)

! key cal constant

3460 M_adj=Cal_real(4)

! key cal constant

3470 N_base=Cal_real(5)

! -base

3480 !
3490 !Check that cal constants are reasonable--if not, use nominal values
3500 IF (M_plus<-.005 OR M_plus>-.003) THEN M_plus=-3.834E-3
3510 IF (M_minus<-.005 OR M_minus>-.003) THEN M_minus=-3.834E-3
3520 IF (M_adj<-.0012 OR M_adj>-.0009) THEN M_adj=-.001021
3530 IF (P_base<3180 OR P_base>3889) THEN P_base=3535
3540 IF (N_base<10 OR N_base>245) THEN N_base=128
3550 !

(Continued on next page)