Test 2-6: AC Flatness - 10 MHz Filter

Description

The purpose of this test is to verify that the AFG meets its specifications for
AC flatness with the 10 MHz filter enabled.

Equipment Setup

Connect equipment as shown in Figure 2-6

Set DMM to ACV, autorange

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Set up the AFG to output a 24 dBm sinewave with the 10 MHz filter
enabled:

VOLT 24DBM;

Set amplitude

:OUTP:LOAD 50

50 ohm load

OUTP:FILT:FREQ 10MHZ

10 MHz filter

OUTP:FILT ON

Enable filter

INIT:IMM

Initiate waveform

3.

Set AFG output to the reference frequency (1 kHz):

FREQ 1000

Set frequency

4.

Measure the amplitude with the DMM, convert the reading to dBm,
and note the reading for future reference:

Reference Level

(

dBm

)

=

20

×

log

ï

Reading

(

volts

)ï+

13.0103

5.

Set the AFG to the crossover frequency (lowest frequency that the
Power Meter can measure):

FREQ 1E5

Set frequency

6.

Measure the amplitude with the DMM and note the reading for future
reference.

52 Verification Tests

Agilent E1445A Service Manual