Test 2-9: Total Harmonic Distortion

Description

The purpose of this test is to verify that the AFG meets its specifications for
sine wave total harmonic distortion (THD).

Equipment Setup

Connect equipment as shown in Figure 2-10

Set Spectrum Analyzer to:

Ref Level = 25 dBm
Freq Span = 1 kHz
Resolution BW = 30 Hz
Video BW = 30 Hz

NOTE

These are recommended settings only. Adjust your Spectrum Analyzer as
necessary.

Figure 2-10. Equipment Setup for Test 2-9 and Test 2-10