Test 2-3: DC Offset

Description

The purpose of this test is to verify that the AFG meets its specifications for
DC offset accuracy.

Equipment Setup

Connect equipment as shown in Figure 2-5

Set DMM to DCV, autorange

Test Procedure

1.

Reset the AFG:

*RST;*CLS

Reset AFG and clear
status registers

2.

Delete all sequences and segments from memory:

LIST:SSEQ:DEL:ALL

Delete all sequences

LIST:SEGM:DEL:ALL

Delete all segments

3.

Create a user-defined waveform made up of zeros:

LIST:SEGM:SEL ZEROS

Select segment name

LIST:SEGM:DEF 8

# of segment points

LIST:SEGM:VOLT 0,0,0,0,0,0,0,0

Segment list

LIST:SSEQ:SEL DC_ZEROS

Select sequence name

LIST:SSEQ:DEF 1

# of segments

LIST:SSEQ:SEQ ZEROS

Sequence list

4.

Set up the AFG to output the waveform defined above:

ROSC:SOUR CLK10;

Select 10 MHz clock

:OUTP:LOAD INF;

Infinite load

:VOLT MAX;

Set amplitude

:FUNC USER

Select user waveform

FUNC:USER DC_ZEROS

Select sequence

INIT:IMM

Initiate waveform