Test 2-4: AC Accuracy (cont’d)

Test Procedure
(cont’d)

4.

Set the AFG output amplitude:

VOLT <amplitude>VRMS

Set amplitude

where <amplitude> is the value specified in Table 2-4.

5.

Trigger the DMM and record the reading.

Table 2-4. AC Accuracy Test Points

Amplitude

(volts rms)

Filter

Test

Limits

±

(dB)

7.2390
6.4500
5.7500
4.5660
2.8818
1.4444
0.2290
7.2390
7.2390

None
None
None
None
None
None
None

250 kHz

10 MHz

0.10
0.15
0.15
0.15
0.15
0.15
0.15
0.10
0.10

46 Verification Tests